Service Guide E8364-90038
3- 47
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362C, E8363C, E836C
Adjustments
Receiver Calibration Adjustment
The receiver calibration is used to adjust the network analyzer receivers for a flat response across its full
frequency range:
1. A power meter/sensor is connected to Port 1, as shown in
Figure 3-22
, to establish a reference for
flatness.
2. A cable is inserted between the power sensor and the test port, as shown in
Figure 3-23
, to establish a
reference for the cable.
3. The same cable is connected between the two test ports, as shown in
Figure 3-24
, and a signal from Port
1 is used to adjust the “B” receiver at Port 2.
The adjustment is repeated using a signal from Port 2 to adjust the “A” receiver at Port 1.
Data obtained during this adjustment are stored in the mxcalfile_pxx files on the hard disk drive (where xx is
a 1 or 2 digit number). The data are used in subsequent measurements.
If the hard disk drive is replaced, these mxcalfile_pxx files will be lost. Therefore, they should be backed up
(saved on a USB drive) so that they can be restored. If using multiple disk drives (e.g. classified and general
usage), then these files must be replaced on each individual disk drive.
These files can be recreated by performing another receiver calibration adjustment.
Equipment Used for the Receiver Calibration Adjustment
Procedure
1. Connect the equipment as shown in
Figure 3-22
. Connect a GPIB cable between the network analyzer
and the power meter.
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
Receiver Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
, and then follow the instructions as they are displayed.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model Part
Number
Power meter
All
E4418B/E4419B
E4418A/E4419A
Power sensor, 3.5 mm
E8362C
E4413A
8485A
Adapter, 3.5 mm (f) to 3.5 mm (f)
E8362C
83059B
85052-60012
RF Cable, 3.5 mm (f) to 3.5 mm (f)
E8362C
85131C
85131E
Power sensor, 2.4 mm
E8363C, E8364C
8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
E8363C, E8364C
11900B
85056-60007
RF Cable, 2.4 mm (f) to 2.4 mm (f)
E8363C, E8364C
85133C
85133E
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...