Installation Note E8364-90011
21
Step 11. Enable Option UNL
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse is required.
Enable Option UNL
1. On the analyzer’s
System
menu, point to
Service
, and then click
Option Enable
.
2. In the
Select Desired Option
list, click
UNL - Bias Tees w/Atten
.
3. Click
Enable
.
4. Click
Yes
in answer to the displayed question in the
Restart Analyzer?
box.
5. When the installation is complete, click
Exit
.
Verify that Option UNL is Enabled
1. On the analyzer’s
Help
menu, click
About Network Analyzer
.
2. Verify that “UNL” is listed after “Options:” in the display.
3. Click
OK
when done.
NOTE
If Option UNL has not been enabled, perform
again. If the
option is still not enabled, contact Agilent Technologies. Refer to
Assistance from Agilent” on page 3
.