Step 6.
Reinstall the Front Panel Assembly
CAUTION
Use a 5/16 in torque wrench (set to 10 in-lbs) on all SMA cable connections in
this step of the procedure.
CAUTION
Before installing the front panel assembly onto the analyzer, lift and support the
front of the analyzer chassis.
Refer to
1. Connect the front panel interface ribbon cable (item
①
).
2. Slide the front panel over the test port connectors being careful to align the power switch
and floppy disk drive to their corresponding front panel cutouts. Ensure that the ribbon
cable
①
is located below the fan to prevent it from being damaged by the fan blades.
3. With a T-10 TORX driver, install the six screws (item
②
) in the sides of the frame.
4. Install the semirigid jumpers (item
③
) on the front panel.
Figure 7