Power Meter Remote Operation
Status Reporting
1-92
EPM-P Series Power Meters Programming Guide
Device Status Register
The device status register set contains bits which give device dependent
information.
The following bits in these registers are used by the power meter:
Bit
Number
Decimal
Weight
Definition
0
-
Not used
1
2
Channel A sensor connected
2
4
Channel B sensor connected
1
1. E4417A only
3
8
Channel A sensor error
4
16
Channel B sensor error
1
5
32
Channel A sensor Front/Rear
6
64
Channel B sensor Front/Rear
1
14
16384
Front Panel key press