Power Meter Remote Operation
Status Reporting
EPM-P Series Power Meters Programming Guide
1-95
Examples
This example program uses the
*OPC?
command to determine when the
power meter has finished calibrating.
CAL:AUTO ONCE
*OPC?
MEAS:POW:AC?
This example program, in HP Basic, uses the
*OPC
command and serial
poll to determine when the power meter has finished calibrating. The
advantage to using this method over the
*OPC?
command is that the
computer can perform other operations while it is waiting for the power
meter to finish calibrating.
10 ASSIGN @Power TO 713
20 OUTPUT @Power;“*CLS”
30 OUTPUT @Power;“*ESE 1”
40 OUTPUT @Power;“CAL:AUTO ONCE;*OPC”
50 WHILE NOT BIT(SPOLL(@Power),5)
60 !(Computer carries out other operations here)
70 END WHILE
80 OUTPUT @Power;“MEAS:POW:AC?”
90 ENTER @Power;Result
100 PRINT Result
110 END