Measurement Commands
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RATio:RELative [<expected_value>[,<resolution>[,<source
list>]]]
2-22
EPM-P Series Power Meters Programming Guide
Example
source list
This channel list specifies the
channels used to calculate the
ratio.
If unspecified and the current
window setup is a ratio
measurement then this ratio setup
is used, otherwise it defaults to
channel A/B (E4417A) or A/A
(E4416A).
(@1),(@2)
3
(@2),(@1)
3
(@1),(@1)
(@2),(@2)
3
1. The mnemonic
DEF
means DEFault. This is not equivalent to the DEFault
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right. The
parameter DEFault is used as a place holder. Specifying
DEF
leaves the parameter
value unchanged.
2. When the measurement result is linear this parameter represents the number of
significant digits. When the measurement result is logarithmic 1 to 4 represents a
resolution of 1, 0.1, 0.01 and 0.001 respectively.
3. E4417A only.
CONF1:RAT:REL DEF,1,(@1),(@2)
This command configures the
upper window to make a ratio
measurement of channel A over
channel B with relative mode on,
using the current sensor range and
a resolution setting of 1 on both
channels.
Item
Description/Default
Range of Values