Measurement Commands
READ[1]|2|3|4 Commands
EPM-P Series Power Meters Programming Guide
2-39
READ[1]|2|3|4 Commands
The
READ?
commands are most commonly used with the
CONFigure
command to cause a new power measurement to be taken and the result
returned to the output buffer. The format of the result is set by
FORM[:READ][:DATA]
. Refer to chapter 6, Format, for further
information.
.
•
For the E4416A the
READ?
query is equivalent to:
ABORt
INITiate
FETCh?
•
For the E4417A carrying out a single channel measurement the
READ?
queries are equivalent to:
ABORt1
INITiate1
FETCh1?
or
ABORt2
INITiate2
FETCh2?
•
For the E4417A carrying out a difference measurement the
READ:DIFFerence?
queries are equivalent to:
ABORt1and
ABORt2
INITiate1
INITiate2
FETCh:DIFFerence?
•
For the E4417A carrying out a ratio measurement the
READ:RATio?
queries are equivalent to:
ABORt1
ABORt2
INITiate1
INITiate2
FETCh:RATio?