CALCulate Subsystem
CALCulate[1]|2|3|4LIMit:FAIL?
3-16
EPM-P Series Power Meters Programming Guide
CALCulate[1]|2|3|4LIMit:FAIL?
This query enters a
1
or
0
into the output buffer indicating whether there
have been any limit failures for the specified window. A limit failure is
defined as
CALC[1]|2|3|4:LIMit:FCO?
being non-zero. The FCO (fail
counter) can be zeroed using the
CALC[1]|2|3|4:LIMit:CLEar
command.
•
1
is returned when one or more limit failures have occurred.
•
0
is returned when no limit failures have occurred.
Syntax
Example
Reset Condition
On reset, the buffer is set to zero for both upper and lower window
measurements.
CALC1:LIM:FAIL?
This command queries if there
have been any limit failures on the
upper window/upper
measurement.
:LIM
:FAIL
?
CALC
1
2
3
4