CALCulate Subsystem
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA] <numeric_value>
3-20
EPM-P Series Power Meters Programming Guide
Parameters
Example
Reset Condition
On reset, both measurements of the upper and lower windows are set to
-90.00 dBm or -90 dB (
DEF
).
Query
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA]? [MIN|MAX]
The query returns the current setting of the lower limit or the values
associated with
MIN
and
MAX
for the specified window.
Query Example
Item
Description/Default
Range of Values
numeric_value
A numeric value for the lower test
limit:
•
DEF:
the default is -90.00 dBm
or -90 db.
•
MIN
: -150 dBm or -180 dB.
•
MAX:
+230 dBm or +200 dB.
-150 to +230 dBm or
-180 to +200 dB
DEF
MIN
MAX
CALC2:LIM:LOW:DATA 0.1
This command enters a lower limit
for the lower window/upper
measurement depending on the
window’s units as follows:
dBm = 0.1 dBm
W = 100 mW
dB = 0.1 dB
% = 0.1 %
CALC2:LIM:LOW:DATA?
This command queries the lower
limit set for the lower window
upper measurement.