CALCulate Subsystem
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA] <numeric_value>
EPM-P Series Power Meters Programming Guide
3-21
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA] <numeric_value>
This command enters a value for the upper test limit for the specified
window/measurement used in the
CALCulate[1]|2|3|4:LIMit
:FAIL?
test. The units used are dependent on the current setting of
UNIT:POWer
and
CALCulate:RELative:STATe
as shown in Table 3-2.
When the measured power is greater than the value specified in
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA]
,
CALCulate[1]|2|3|4:LIMit:FAIL?
reports a fail. When the measured
level is less than or equal to the limit, a fail is not reported.
Table 3-2: Measurement Units
Syntax
Measurement
Mode
Measurement
Type
CALC:REL:STAT OFF
CALC:REL:STAT ON
Linear
Log
Linear
Log
Single
Channel
Avg, Pk
Watt
dBm
%
dB
Pk-Avg
%
dB
%
dB
Ratio
Avg, Pk, Pk-Avg
%
dB
%
dB
Difference
Avg, Pk
Watt
dBm
%
dB
Pk-Avg
%
dB
%
dB
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:LIM
:UPP
:DATA
CALC
1
2
3
4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com