DISPlay Subsystem
DISPlay:CONTrast <numeric_value>
EPM-P Series Power Meters Programming Guide
5-3
DISPlay:CONTrast <numeric_value>
This command controls the display contrast:
• A contrast of
0
represents a minimum contrast.
• A contrast of
1
represents a maximum contrast.
When the supply power is cycled off then on the contrast sets to the
factory default for that particular power meter.
Syntax
Parameters
Example
Item
Description/Default
Range of Values
numeric_value
A numeric value for the contrast level.
•
DEF:
the factory default value
which can vary between power
meters.
•
MIN:
0.
•
MAX
: 1.
Units are resolved to 0.01.
0
to
1
DEF
MIN
MAX
DISP:CONT 0.75
This command sets the display
contrast to 0.75.
DISP
:CONT
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com