DISPlay Subsystem
DISPlay[:WINDow[1]|2][:NUMeric[1]|2]:RESolution <numeric_value>
EPM-P Series Power Meters Programming Guide
5-21
DISPlay[:WINDow[1]|2][:NUMeric[1]|2]:RESolution <numeric_value>
This command sets the resolution of the measurement result in the
specified window.
Syntax
Parameters
Example
Reset Condition
On reset, the resolution is set to 3.
Item
Description/Default
Range of Values
numeric_value
A numeric value for the window
resolution:
•
DEF
: 3
•
MIN
: 1
•
MAX
: 4
1 to 4
DEF
MIN
MAX
DISP:WIND2:RES 4
This command sets the lower
window’s resolution to four
significant digits if the
measurement result is linear, or to
0.001 if the measurement result is
logarithmic.
:RES
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:NUM
:WIND
1
2
DISP