MEMory Subsystem
MEMory Subsystem
7-2
EPM-P Series Power Meters Programming Guide
MEMory Subsystem
The
MEMory
command subsystem is used to:
•
Edit and review sensor calibration tables (8480 Series power
sensors and N8480 Series sensors with Option CFT only).
•
Store sensor calibration tables (8480 Series power sensors and
N8480 Series sensors with Option CFT only).
•
Edit and review sensor frequency dependent offset tables.
•
Store sensor frequency dependent offset tables.
•
Edit and review sensor save/recall registers.
Stored tables remain in the power meter’s memory during power down.
The power meter is capable of storing 20 sensor calibration tables and 10
frequency dependent offset tables of 80 frequency points each.
Note
The
MEMory
subsystem is not used for E-Series power sensors and
N8480 Series power sensors (excluding Option CFT) calibration
tables which are automatically downloaded to the power meter
and cannot be reviewed or edited.
Keyword
Parameter Form
Notes
Page
MEMory
:CATalog
[:ALL]?
[query only]
page 7-5
:STATe?
[query only]
page 7-7
:TABLe?
[query only]
page 7-8
:CLEar
[:NAME]
<character_data>
[no query],
[non-SCPI]}
page 7-11
:TABLe
[no query]
page 7-12
:FREE
[:ALL]?
[query only]
page 7-14
:STATe?
[query only]
page 7-15
:TABLe?
[query only]
page 7-16
:NSTates?
[query only]
page 7-17
:STATe
:CATalog?
[query only]
page 7-19
:DEFine
<character_data>
[,<numeric_value>]
[non-SCPI]
page 7-20
:TABLe
:FREQuency
<numeric_value>
[,<numeric_value>]
page 7-23
:POINts?
[query only]
page 7-26