MEMory Subsystem
MEMory:CLEar[:NAME] <character_data>
EPM-P Series Power Meters Programming Guide
7-11
MEMory:CLEar[:NAME] <character_data>
This command clears the contents of a specified sensor calibration table
(8480 Series power sensors and N8480 Series sensors with Option CFT
only), frequency dependent offset table, or save/recall register.
Although the table remains, a
MEMory:TABLe:FREQuency|GAIN:POINts?
query returns a
0
as there
are no contents in the table.
For sensor calibration tables and frequency dependent offset tables, this
command is an alternative form of the
MEMory:CLEar:TABLE
command,
the only difference being the method in which the table is selected.
Note
The contents cleared using this command are non-recoverable.
Syntax
Parameters
Example
Error Messages
If the table or save/recall register name does not exist, error -224, “Illegal
parameter value” occurs.
Item
Description/Default
Range of Values
character_data
Contains an existing table name or
save/recall register.
Any existing table
name or save/recall
register.
MEM:CLE "HP8485A"
This command clears the contents
of sensor calibration table
HP 8485A
MEM
:CLE
:NAME
Space