MEMory Subsystem
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
7-24
EPM-P Series Power Meters Programming Guide
Syntax
Parameters
Example
Query
MEMory:TABLe:FREQuency?
The query returns a list of frequency points for the table currently
selected. The frequencies are returned in Hz.
Query Example
Item
Description/Default
Range of Values
numeric_value
A numeric value for the frequency. The
default units are Hz.
1 kHz to
999.9 GHz
1,
2
1. The following measurement units can be used:
•Hz
•kHz (10
3
)
•MHz (10
6
)
•GHz (10
9
)
2. All frequencies are truncated to a multiple of 1 kHz.
MEM:TABL:FREQ 200kHz,600kHz
This command enters
frequencies of 200 kHz and
600 kHz into the currently
selected table.
MEM:TABL:FREQ?
This command queries the
frequency points in the currently
selected table.
MEM
:TABL
:FREQ
numeric_value
Space
,
?