SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:SDETect <boolean>
EPM-P Series Power Meters Programming Guide
9-11
[SENSe[1]]|SENSe2:AVERage:SDETect <boolean>
This command enables and disables step detection. In
AUTO
filter mode,
the average of the last four values entered into the filter is compared to
the average of the entire filter. If the difference between the two averages
is greater than 12.5%, the digital filter is cleared. The filter then starts
storing new measurements. This feature is known as step detection and
shortens the settling time of the filter when the input power changes
substantially.
Note
Step detection is automatically disabled under the following
circumstances:
•
TRIG:DEL:AUTO
is
ON
and the trigger mode is set to free run.
or,
•
Auto-averaging is disabled.
Under the above circumstances the value of
SENS:AVER:SDET
is
ignored. Note also that
SENS:AVER:SDET
is not set by the instrument
(that is,
SENS:AVER:SDET
will retain its current setting which may
indicate that step detection is
ON
).
Syntax
Example
Reset Condition
On reset, step detection is enabled.
SENS:AVER:SDET OFF
This command disables step
detection.
SENS
1
AVER
SENS2
:
:SDET
0|OFF