SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <boolean>
EPM-P Series Power Meters Programming Guide
9-33
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <boolean>
This command is used to enable and disable the pulse power
measurement feature.
The
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3
command
should be used to enter the duty cycle of the signal you want to measure.
Syntax
Example
Reset Condition
On reset, the pulse power measurement feature is disabled.
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe?
The query enters a
1
or
0
into the output buffer indicating the status of
the pulse power measurement feature.
•
1
is returned when the pulse power measurement feature is
enabled.
•
0
is returned when the pulse power measurement feature is
disabled.
Query Example
CORR:DCYC:STAT 1
This command enables the pulse
measurement feature on
channel A.
CORR:GAIN3:STAT?
This command queries whether
the pulse measurement feature is
on or off.
SENS
1
:DCYC
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?
:GAIN3