STATus Subsystem
STATus:OPERation
EPM-P Series Power Meters Programming Guide
10-11
STATus:OPERation
The operation status register set contains conditions which are a part of
the operation of the power meter as a whole.
The following bits in these registers are used by the power meter:
Syntax
Bit
Number
Decimal
Weight
Definition
0
1
CALibrating Summary
1 - 3
-
Not used
4
16
MEASuring Summary
5
32
Waiting for TRIGger Summary
6 - 9
-
Not used
10
1024
SENSe Summary
11
2048
Lower Limit Fail Summary
12
4096
Upper Limit Fail Summary
13 to 15
-
Not used (bit 15 always 0)