STATus Subsystem
STATus:QUEStionable
10-20
EPM-P Series Power Meters Programming Guide
STATus:QUEStionable
The questionable register set contains bits which give an indication of the
quality of various aspects of signals processed by the power meter as a
whole.
The following bits in these registers are used by the power meter:
Bit 3 is set by the logical OR outputs of the
STATus:QUEStionable:POWer:SUMMary
register set.
Bit 8 is set by the logical OR outputs of the
STATus:QUEStionable:CALibration:SUMMary
register set.
Bit 9 is set if power-on self-test fails, and cleared if it passes.
Syntax
Bit
Number
Decimal
Weight
Definition
0 to 2
-
Not used
3
8
POWer Summary
4 to 7
-
Not used
8
256
CALibration Summary
9
512
Power On Self Test
10 to 15
-
Not Used (bit 15 always 0)
STAT