SYSTem Subsystem
SYSTem:PRESet <character_data>
Agilent E4416A/E4417A Programming Guide
11-39
Table 11-5:
EDGE
: Power Meter Presets
Command
Setting
Comments
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX]
+900.000 MHz
Frequency setting
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
NORM
Measurement mode
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
E9321A/25A: HIGH
E9322A/26A: MED
E9323A/27A: LOW
Sensor video bandwidth
Gate setup
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
Gate 1: 20 us
Gates 2 - 4: 0
Delay between trigger point and
time gated period.
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Gate 1: 520 us
Gates 2 - 4: 0
Length of time gated period for
time gated measurements
Trigger setup
TRIGger[:SEQuence[1]|2]:SOUR
INT1
Trigger source setup and
acquisition mode continuous
triggering
INIT:CONT
ON
TRIG[:SEQ]:LEV:AUTO
OFF
Disable automatic setting of the
trigger level
TRIG[:SEQ]:LEV
-20 dBm
Power level
TRIG[:SEQ]:SLOP
POS
Trigger event recognized on the
rising edge of a signal
TRIG[:SEQ]:DEL
0
Delay between recognition of
trigger event and start of a
measurement
TRIG[:SEQ]:HOLD
4275 us
Trigger holdoff
Step detection
[SENSe[1]]|SENS2:AVER:SDET
1
Step detection enabled
Trace setup
DISP[:WIND[1]|2]:TRAC:LOW
+20 dBm
Maximum power
DISP[:WIND[1]|2]:TRAC:UPP
-35 dBm
Minimum power
[SENS[1]]|SENS2:TRAC:OFFS
:TIME <numeric_value>
-40 us
Delay between delayed trigger
point and the start of the trace
[SENS[1]]|SENS2:TRAC:TIME
<numeric_value>
700 us
Length of the trace
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com