TRIGger Subsystem
TRIGger[:SEQuence]:HYSTeresis <numeric_value>
13-24
EPM-P Series Power Meters Programming Guide
TRIGger[:SEQuence]:HYSTeresis <numeric_value>
This command sets:
•
How far a signal must fall below
TRIG:LEVel
before a rising edge
can be detected.
•
How far a signal must rise above
TRIG:LEVel
before a falling
edge can be detected.
Note
The command is accepted for
TRIGger[:SEQuence[1]]
(channel A)
only, for both single and dual channel power meters.
Syntax
Example
Parameters
Reset Condition
On reset the value is set to 0 dB.
TRIG:SEQ:HYST 0.1
This command sets the value to
2 dB for channel A.
Item
Description/Default
Range of Values
numeric_value
How far a signal must fall/rise before
a rising or falling edge can be
detected.
DEF:
the default value is 0 dB.
Units are resolved to 0.05 dB.
0 to 3 dB
DEF
:HYST
Space
numeric_value
?
DEF
TRIG
:SEQ
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com