SERVice Subsystem
SERVice:BIST:FPATH[1]|2:MEASure?
15-4
EPM-P Series Power Meters Programming Guide
SERVice:BIST:FPATH[1]|2:MEASure?
This query returns the power measured by the meter as a DC voltage, for
the specified channel. It applies to E9320 Series power sensors only.
Note
Additional hardware is needed to run this command. For further
information refer to the power meter Service Guide.
Syntax
Example
Error Messages
• If no power sensor is connected, error -241 “Hardware missing” occurs.
• If a non E9320 Series power sensor is connected, error -241
“Hardware missing” occurs.
• If
FPATH2
is sent to a single channel meter, error -113
“Undefined header” occurs.
SERV:BIST:FPATH:MEAS?
This command queries DC voltage
measured by the power meter on
channel A.
:BIST
SERV
?
:FPATH
:MEAS
1