SERVice Subsystem
SERVice:SENSor[1]|2:CORRections:STATe <boolean>
15-14
EPM-P Series Power Meters Programming Guide
SERVice:SENSor[1]|2:CORRections:STATe <boolean>
This command enables/disables the voltage to corrected power conversion.
It applies to E9320 Series power sensors only.
Note
Before setting this command to
OFF
, you must set the
INIT:CONF
command to
OFF.
After setting this command to
OFF
, you must only run commands
relating to the gathering of ADC values—for example, the
SERV:SENS:RADC
command.
Syntax
Example
Reset Condition
On reset, the value is set to
OFF
.
Query
SERVice:SENSor[1]|2:CORRections:STATe?
The query enters a
1
or
0
into the output buffer indicating the status of
the voltage to corrected power conversion.
•
1
is returned when voltage to corrected power conversion is enabled.
•
0
is returned when voltage to corrected power conversion is disabled.
SERV:SENS2:CORR:STAT ON
This command enables the voltage
to corrected power conversion for
channel B.
0|OFF
1|ON
Space
?
:STAT
:CORR
SERV
1
2
:SENS