Calibration Factor Block Layout
Calibration Factor Block Layout
A-6
Agilent E4416A/E4417A Programming Guide
Cal factor (low power)
1
2
-
2.14
0.25 to 3
None
Power (in watts) is divided by
this value.
Cal factor (high power)
1
2
-
2.14
0.25 to 3
None
Power (in watts) is divided by
this value.
These table entries are repeated as shown for each frequency point
Frequency (point ‘N’)
4
-
32 bit fixed
0 to Fhpb* (2^32)
None
Fhbp = Freq Hz per bit
Cal factor (low power)
1
2
-
2.14
0.25 to 3
None
Power (in watts) is divided by
this value.
Cal factor (high power)
1
2
-
2.14
0.25 to 3
None
Power (in watts) is divided by
this value.
Table Size:
-
See note
1
The table size is dependent on
the number of freq. points.
1.
Corrections are applied in power for E4410, E9300 and N8480 Series sensors
(excluding Option CFT).
2.
Corrections are applied in voltage versus ADC reading for E9320 Series sensors. This
format also requires only one correction factor across all power levels.
3.
The block layout shown for E9320 Series sensors exists in two separate EEPROM
locations. One location contains the calibration factor data for the average path and
the other contains the calibration factor data for the peak path. These EEPROM
blocks are accessed using the
SERV:SENS:CALFactor
and
SERV:SENS:PCALfactor
commands respectively.
N8480 Series Sensors:
Calibration Factor
Block Layout
No.
Bytes
Contents
Data Format
Data Range
Units
Notes