56
Chapter 1
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
the corresponding bit in the Questionable Calibration Extended Needed Event register
when the condition register bit has a positive transition (0 to 1). The variable <integer> is
the sum of the decimal values of the bits that you want to enable.
Questionable Frequency Register
Questionable Frequency Condition
This query returns the decimal value of the sum of the
bits in the Questionable Frequency Condition register.
NOTE
The data in this register is continuously updated and reflects the
current conditions.
Questionable Frequency Enable
This command determines which bits in the Questionable
Frequency Condition Register will set bits in the Questionable Frequency Event register,
which also sets the Frequency Summary bit (bit 5) in the Questionable Register. The
variable <integer> is the sum of the decimal values of the bits you want to enable.
Mode
All
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition
<integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition?
Example
STAT:QUES:CAL:EXT:NEED:PTR 2 Align EMI conducted is required.
Preset
32767
Min
0
Max
32767
SCPI Status
Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:FREQuency:CONDition?
Example
STAT:QUES:FREQ:COND?
Preset
0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:FREQuency:ENABle <integer>
:STATus:QUEStionable:FREQuency:ENABle?