Chapter 1
63
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
current conditions.
Questionable Integrity Uncalibrated Enable
This command determines which bits in the
Questionable Integrity Uncalibrated Condition Register will set bits in the Questionable
Integrity Uncalibrated Event register, which also sets the Data Uncalibrated Summary bit
(bit 3) in the Questionable Integrity Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.
Questionable Integrity Uncalibrated Event Query
This query returns the decimal value of
the sum of the bits in the Questionable Integrity Uncalibrated Event register.
NOTE
The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Mode
All
Remote
Command
:STATus:QUEStionable:INTegrity:UNCalibrated:CONDition?
Example
STAT:QUES:INT:UNC:COND?
Preset
0
SCPI Status
Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
?
Example
STAT:QUES:INT:UNC:ENAB 1 Oversweep (Meas Uncal) will
be reported to the Integrity Summary of the Status
Questionable register.
Preset
32767
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?