68
Chapter 1
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
NOTE
The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Questionable Temperature Negative Transition
This command determines which bits in the
Questionable Temperature Condition register will set the corresponding bit in the
Questionable Temperature Event register when the condition register bit has a negative
transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Questionable Temperature Positive Transition
This command determines which bits in the
Questionable Temperature Condition register will set the corresponding bit in the
Questionable Temperature Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature[:EVENt]?
Example
STAT:QUES:TEMP?
Preset
0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:NTRansition
<integer>
:STATus:QUEStionable:TEMPerature:NTRansition?
Example
STAT:QUES:TEMP:NTR 1 Reference Oscillator Oven not cold
will be reported to the Temperature Summary of the Status
Questionable register.
Preset
0
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode