Chapter 3
87
Measurement Synchronization with LXI-B
Peer to Peer Messaging
Peer to Peer Messaging
MXA Instrument Events
The MXA and MXG are both capable of sending and receiving LXI
specific LAN packets. The packets are configurable, and may be sent
when various instrument events occur during a measurement or state
recall. Each instrument event has an associated sense of 0 or 1 to
indicate whether or not the event is active. The instrument events that
can cause an MXA to send an LXI LAN packet are summarized in the
table below.
NOTE
As of this product firmware release in June 2008, LXI Class B
Compliance Tests, using the new IEEE 1588-2008 Precision Time
Protocol (PTP), were not available. This product provides the features of
an LXI Class B instrument by adding LAN Triggering and Time
Synchronization to its LXI Class C compliance. References to LXI Class
B in the document and in the instrument menus and web pages do not
imply LXI Class B compliance. This product will apply for LXI Class B
Compliance when the LXI Consortium has approved its specification
changes and can provide certified testing.
Outputting LAN Events
The MXA can be configured to send LXI Output LAN Events using
either the front panel or SCPI by following the procedures below.
Instrument Event
Description
OperationComplete
Indicates whether or not a measurement operation is underway. For
example, a single sweep with averaging turned off will have
OperationComplete=1 before the sweep, OperationComplete=0
during the sweep, and OperationComplete=1 after the sweep. If
averaging is turned on, however, OperationComplete will remain low
until the number of sweeps required for the full average have been
executed.
WaitingForTrigger
Indicates whether or not the instrument is waiting for a trigger
event before a measurement begins. WaitingForTrigger will
transition high before every sweep and will return low after the
trigger occurs.
Sweeping
Indicates whether or not the instrument is actively sweeping the LO.
Measuring
Indicates whether or not the instrument is actively measuring its
input signal.
Recalling
Indicates whether or not the instrument is recalling a state.