130
N9912A FieldFox User’s Guide
Display Resolution
You can display either one digit or two digits after the decimal point for both
magnitude and phase readings.
This setting survives a Preset and Mode Preset.
How to set Resolution
Press
Scale/Amptd
Then
Resolution
Then choose from:
o
1
0.0 (default)
o
2
0.00
IF Bandwidth
Reducing the IF receiver bandwidth reduces the effect of random noise on a
measurement. However, narrower IF bandwidths also cause longer measurement
times.
For highest accuracy, perform calibration and Zero AFTER changing the IF
Bandwidth. Learn about Zeroing on page 131.
How to set IF BW
Press
BW 2
.
Then
IF BW
Then choose from the following:
300 Hz | 1 kHz | 3 kHz | 10 kHz | 30 kHz
Output Power
Set the power level out of the FieldFox to either High power or Low
power. Generally, the high power level (default setting) is more accurate because
the receivers are measuring farther from the noise floor. However, some devices
are sensitive to high power levels. Use the Low power setting when measuring
amplifiers.
How to set Output Power
Press
Meas Setup 4
then
More
Then
Output Power
Then choose from the following. The underlined value is the current setting.
o
High
Sets output power to the maximum achievable power at the selected
frequency.
o
Low
Sets output power to approximately –45 dBm.
Averaging
Averaging helps to reduce the effects of random noise on a measurement. The
FieldFox computes the displayed value over several consecutive readings.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...