VVM (Vector Voltmeter) Mode – Option 308
131
Average Count determines the number of readings to average. The higher the
average count, the greater the amount of noise reduction.
An average counter is shown in the left edge of the screen as
Avg N
. This shows
the number of previous readings that have been averaged together. When the
counter reaches the specified count, then a ‘running average’ of the last
N
readings is displayed.
Average Count = 1 means there is NO averaging.
This setting can be changed at any time without affecting calibration accuracy.
How to set Averaging
Press
Meas Setup 4
Then
More
Then
Average N
where
N
is the current count setting.
Enter a value using the numeric keypad. Enter
1
for NO averaging.
Press
Enter
.
While Averaging is in process, press
Meas Setup 4
, then
More
, then
Restart
to restart the averaging.
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
Single
button is pressed. Use
Single
to conserve battery
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
Press
Sweep 3
.
Then choose one of the following:
o
Single
Automatically sets Continuous OFF and causes FieldFox to make
ONE measurement, then hold for the next Single key press. When a data
trace is displayed, the entire trace is measured, then holds. The
Hold
annotation changes to an arrow
-->
while the measurement occurs.
o
Continuous
Makes continuous measurements. This is the typical setting
when battery power is not critical.
You can also use
Run / Hold +/-
to toggle between Single and Continuous.
Zeroing
When performing a Cable Trimming operation, one cable is compared to another.
After measuring the first, shorter cable, pressing Zero performs the same
operation as Data->Memory, then Data/Mem, also known as Normalization. All
subsequent electrical length measurements are relative to the stored
measurement.
To make an accurate measurement, set
Zero OFF
before calibrating.
Zero after changing the IF Bandwidth.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...