26
N9912A FieldFox User’s Guide
Enter a value using the numeric keypad. Enter 1 for NO averaging.
Press
Enter
.
While Trace Averaging is in process, press
Sweep 3
then
Restart
to restart
the averaging at 1.
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
Single
button is pressed. Use
Single
to conserve battery
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
Press
Sweep 3
.
Then choose one of the following:
o
Single
Automatically sets Continuous OFF and causes FieldFox to make
ONE measurement, then hold for the next Single key press. When a data
trace is displayed, the entire trace is measured, then holds. The
Hold
annotation changes to an arrow
-->
while the measurement occurs.
o
Continuous
Makes continuous measurements. This is the typical setting
when battery power is not critical.
You can also use
Run / Hold +/-
to toggle between Single and Continuous.
Resolution (Number of Data Points)
Data points are individual measurements that are made and plotted across the X-
axis to form a trace. Select more data points to increase measurement resolution.
However, more data points require more time to complete an entire
measurement sweep.
When the Resolution is changed after a calibration is performed, the cal becomes
interpolated. Learn more on page 65.
How to set Resolution
Press
Sweep 3
.
Then
Resolution
.
Then choose one of the following:
101 | 201 | 401 | 601 | 801 | 1001 |1601 | 4001 | 10001
Using SCPI, Resolution can be set to any number of points between 3 and
10001. See the Programming Guide at
www.agilent.com/find/fieldsfoxsupport
Sweep Time
The fastest possible sweep time is always used as the default setting. Use the
Min
Swp Time
setting to slow the sweep time when measuring long lengths of cable.
Learn more in the
Supplemental Online Help:
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...