Increase Dynamic Range
Dynamic range is the difference between maximum input power to the FieldFox
receiver (without compressing the receiver), and the minimum measurable
power (noise floor). Measurement accuracy is increased when the DUT response
is at least 10 dB above the noise floor. For a measurement to be valid, input
signals must be within these boundaries.
A common application that requires a wide dynamic range is measuring the
rejection band of a base station filter.
The following steps MAY help you increase the dynamic range of your
NA Mode
measurement. The results you see will depend on the performance of your DUT.
1. Set Start and Stop frequency.
2. Press
Meas 4
then
Output Power High
3. Lower the IFBW. Press
BW 2
then
IF BW
then
300 Hz
4. Increase Averaging. Press
BW 2
then
Average
then enter 10.
5. Select S21.
6. Perform a Normalization Cal. Learn how page 64.
7. With RF OUT (port-2) open, press
Trace 6
then
Data->Mem
8. Connect the DUT.
The following image shows S21 measurements of a high-dynamic range filter
using the above procedure.
Blue trace – measurement WITHOUT step 7.
Black trace – measurement WITH step 7.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...