SA (Spectrum Analyzer) Mode
79
How to use the Antenna/Cable Editor
Press
New
then
Yes
to clear all data from the existing Antenna or Cable table
and reset header information to default settings.
Then
Edit Antenna
or
Edit Cable
to open the Antenna / Cable Editor.
Then use the
▲|▼
arrows to select a field,
o
When editing Header information, press
Edit
then modify the selected field
using the FieldFox labeler.
o
When editing Frequency/Value pairs, enter numbers using the numeric
keypad, and a frequency suffix when necessary. Then
Enter
. Positive Values
are interpreted as LOSS for both Cable Loss and Antenna Gain Factor.
Learn more about “How the Freq/Value pairs are applied” below.
Optionally choose from the following:
o
Previous / Next Page
Quickly scrolls through pages of Freq/Value data.
o
Add Data
Add a blank Freq/Value pair to the table,
o
Delete/Clear
then:
o
Delete Line
Remove the selected Freq/Value pair from the table.
o
Clear All
then
Yes
Remove all Freq/Value pairs from the table and resets
header information to default settings.
When the table edits are complete
Press
Done
to closes the Antenna/Cable Editor.
Press
Save Antenna
or
Save Cable
to saves your changes to the specified
Storage Device. Enter a filename using the FieldFox labeler (learn how on page
151). Learn more about Antenna and Cable files below.
Press
Recall Antenna
or
Recall Cable
to load an Antenna or Cable file from
the specified Storage Device.
Storage Device
Changes the device used to save or recall files. Choose from
Internal (default setting), USB (must be connected) or SD card. This is a
different setting from the Save/Recall Storage Device setting.
About the Antenna/Cable header fields
Description:
Enter a description for the antenna or cable correction table. This
is NOT the filename used to save the file.
Antenna Unit:
Units in which the measurement will be displayed. Select
NONE
(default setting) to select from standard SA Units on the Scale menu.
Learn how on page 74.
Choose from:
dB
μV/m
- (dB micro Volts per meter),
dB
μA/m
- (dB micro
Amps per meter),
dBG
(gauss),
dBpT
(tesla).
Freq Interpolation:
Set to LINear and can NOT be edited.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...