SA (Spectrum Analyzer) Mode
83
Source Enable ON OFF
o
ON
Turns ON the internal source.
o
OFF
Turns OFF the internal source.
Source Mode
Choose from the following:
o
Tracking
The internal source sweeps simultaneously (tracks) with the SA
receiver over the displayed frequency range, emulating NA mode
measurements.
Src
is annotated on the screen.
o
CW
The source is set to a CW frequency independent from the SA setting.
Srcw
is annotated on the screen.
o
Coupled CW
The source is set to a CW frequency, but that frequency
“tracks” the SA receiver Center Frequency setting. Therefore, the signal
always appears at center of the display.
Srcp
is annotated on the screen.
Normalize ON OFF
With Source Enable ON and Source Mode =
Tracking
,
Normalize ON stores the current measurement trace into Trace 4 in View state
as a reference trace. All subsequent trace measurements subtract out this
Trace 4 reference trace, and therefore become relative to it. This causes
measurements to appear near the ZERO absolute level on the y-axis (0 dBm if
UNIT is set to dBm).
SrcN
is annotated on the screen. Normalize is
automatically turned OFF when the frequency range or Resolution is changed.
Learn more about Trace Display States on page 89.
Source Power
Sets the source power from -45 dBm to 0 dBm in 1 dB steps.
NOTE
The FieldFox output power is approximate and NOT flat across the frequency
range.
Source CW Freq
Sets the CW frequency that is used when Source Mode is set
to CW.
By default, the source output will turn off momentarily at the end of each SA
sweep. To cause the source to stay ON at the end of each sweep, turn battery
saver OFF. (Learn about Battery Saver on page 164.)
Press
System 7
Then
Preferences
Then
Battery Saver OFF
Resolution Bandwidth (Res BW)
In SA Mode, the Res BW provides the ability to resolve, or see closely spaced
signals. The narrower (lower) the Res BW, the better the spectrum analyzer can
resolve signals. In addition, as the Res BW is narrowed, less noise is measured by
the spectrum analyzer ADC and the noise floor on the display lowers as a result.
This allows low level signals to be seen and measured. However, as the Res BW is
narrowed, the sweep speed becomes slower.
How to set Res BW
Press
BW 2
.
Then
Res BW Auto | Man
.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...