14
N9912A FieldFox User’s Guide
Then
Internal Temperatures
.
The temperature at which the following events occur is the average of the RF1,
RF2, SB1, SB2 temperatures. These temperatures come from internal sensors
embedded within FieldFox.
Temperature Control Mode
At approximately 73
°
C, the FieldFox enters Temperature Control mode by
reducing display intensity and measurement speed. This should decrease the
internal temperature which preserves measurement accuracy and maintains the
long-term reliability of the FieldFox.
When entering Temperature Control mode, save your instrument state and data
that you want to keep.
When the temperature drops to approximately 71
°
C, normal operating settings
are restored.
NOTE
Measurement speed specifications do NOT apply in Temperature Control Mode.
High-Temp Shutdown
In extreme situations, Temperature Control mode may not stop an increase in
the FieldFox internal temperature. At approximately 75
°
C, High-Temperature
Shutdown will engage and turn OFF the FieldFox.
Just prior to shutdown, the FieldFox will display a warning of imminent shut
down.
Avoid Overpowering the FieldFox
The FieldFox can be damaged with too much power or voltage applied.
Exceeding the maximum RF power levels shown below will cause an
ADC Over
Range
message to appear on the screen.
WARNING
Maximum Input Voltages and Power:
RF OUT Connector: ±50 VDC, +23 dBm RF
RF IN: ±50 VDC, +27 dBm RF
DC Input: 19 VDC, 4 ADC
Learn more about Maximum power and voltages in the FieldFox Data Sheet on
page 204.
NOTE
Very often, coaxial cables and antennas build up a static charge, which, if
allowed to discharge by connecting to the FieldFox, may damage the instrument
input circuitry. To avoid such damage, it is recommended to dissipate any static
charges by temporarily attaching a short to the cable or antenna prior to
attaching to the FieldFox.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...