144
N9912A FieldFox User’s Guide
Peak A =
Valid Peak
(Above Threshold and Excursion Settings)
Peak B = Invalid Peak (Below 1 dB Excursion Setting)
Peak C = Invalid Peak (Below –10 dB Threshold Setting)
Marker Functions
Marker functions are used to change FieldFox frequency or distance settings to
those of the active marker.
How to use Marker Functions
Activate the marker to move.
Then press
Mrk ->/Tools
Then
Mkr ->
Then choose one of the following:
o
Mrk -> Center
(Not available in DTF or SA Mode with zero span) The center
frequency of the measurement becomes the value of the active marker. The
frequency span is adjusted accordingly.
o
Mkr -> Delay
(NA Mode ONLY.) The group delay value of the active marker
becomes the electrical delay setting of the trace. Learn how to set group
delay format on page 43. Learn more about electrical delay on page 45.
o
Mkr->RefLev
(SA Mode ONLY) The reference level becomes the magnitude
of the active marker.
o
Mkr->Tune Freq
(SA Mode with Tune & Listen ONLY) The Tune & Listen
Frequency becomes the frequency of the active marker.
The following applies to DTF Measurements ONLY:
o
More
then:
o
Marker=>Start Distance
The Start Distance of the measurement becomes
the value of the active marker. Markers that would no longer appear on the
screen are moved to the new Start Distance.
o
Marker=>Stop Distance
The Stop Distance of the measurement becomes
the value of the active marker. Markers that would no longer appear on the
screen are moved to the new Stop Distance.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...