NA (Network Analyzer) Mode
47
While averaging is in process, press
Sweep 3
, then
Restart
to restart the
averaging at 1.
IF Bandwidth
The FieldFox converts the received signal from its source to a lower intermediate
frequency (IF). The bandwidth of the IF bandpass filter is adjustable. Reducing
the IF receiver bandwidth reduces the effect of random noise on a measurement.
However, narrower IF bandwidths cause longer sweep times.
How to set IF BW
Press
BW 2
.
Then
IF BW
Then choose from the following:
300 Hz | 1 kHz | 3 kHz | 10 kHz | 30 kHz
Smoothing
Trace smoothing averages a number of
adjacent
data points to smooth the peak-
to-peak noise values on a displayed trace. The number of adjacent data points
that are averaged is known as the smoothing aperture
.
Aperture is set by
specifying a percentage of the X-axis span.
Trace smoothing does NOT significantly increase measurement time.
Smoothing is used in Group Delay measurements, although it can be used with
any NA format EXCEPT Polar or Smith Chart. Learn more about NA Mode
formats, including Group Delay, on page 43.
When enabled,
Smo
appears on the FieldFox screen.
How to set Smoothing
Press
BW 2
.
Then
Smoothing ON OFF
Then
Sm. Aperture
and enter a value between 0 and 25 (percent) using the
numeric keypad.
Press
Enter
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
Single
button is pressed. Use
Single
to conserve battery
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
Press
Sweep 3
.
Then choose one of the following:
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...