Output Power
Set the power level out of the FieldFox to High, Low, or manually set power level
to a value between High and Low. Generally, the high power (default) setting is
most accurate because the receivers are measuring farther from the noise floor.
However, some devices are sensitive to high power levels. Use the Low power
setting when measuring amplifiers.
NOTE
To help prevent damage to your DUT, when Output Power is set to High or Low in
NA Mode, it is changed to the same value in CAT Mode. Manual settings in NA
Mode are changed to a Low setting in CAT Mode.
How to set Output Power
Press
Meas Setup 4
.
Then
Output Power
o
High
Sets output power to the maximum achievable power at all displayed
frequencies. Output power is NOT FLAT across the displayed FieldFox
frequency span. Please see the FieldFox Specifications on page 204 for
expected power levels.
o
Low
Sets output power to approximately –45 dBm, FLAT across the
displayed FieldFox frequency span.
o
Man
Set output power to an arbitrary value, FLAT across the displayed
FieldFox frequency span. If flattened power can NOT be achieved, a
warning
message
and beep occurs. To achieve a flattened output power, reduce the
power level or stop frequency.
o
Then press
Power Level
o
Then enter a value using the numeric keypad, the
▲
|
▼
arrows, or the
rotary knob.
o
Press
Enter
.
System Impedance (Z0)
To accurately view data presented in Smith Chart format, first set the System
Impedance.
Learn how to select Smith Chart format on page 43.
Learn how to make 75
Ω
measurements at the FieldFox Supplemental Online
Help:
www.agilent.com/find/fieldsfoxsupport
How to set System Impedance
Press
Meas Setup 4
.
Then
Settings
.
Then scroll to
System Z0
and press
Edit
Then type either
50
or
75
and press
Enter
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...