N9912A FieldFox User’s Guide
Definitions
DUT (Device Under Test)
is the cable, antenna, transmission line, or anything
else that is connected to the FieldFox that is to be measured.
OPEN, SHORT, and LOAD
are calibration standards. These are precision
components that are used during calibration and to terminate a DUT during
some measurements. When an RF signal ‘hits’ these components, the signals are
reflected in a predictable manner.
A SHORT and OPEN standard both cause 100% of an RF signal to be reflected.
The difference between these two standards is what happens to the phase of
the reflected signal, which is beyond the scope of this discussion. Although an
OPEN standard is a precision component, simply leaving
nothing
connected at
the end of a cable can be a reasonable substitute for an OPEN.
A LOAD standard completely absorbs the incident signal and NO signal is
reflected back to the source.
Calibration Reference Plane
is the point at which cal standards are connected
during a calibration. This can be either the FieldFox RF OUT connector, or at the
end of a jumper cable or adapter.
Preset Calibration (also known as CalReady)
Every FieldFox is factory calibrated at the RF Out connector over the entire
frequency range of the FieldFox using a number of data points that allows
reasonable interpolation over the FieldFox frequency range.
This CalReady condition allows you to start using your FieldFox immediately if
you are measuring a device that is directly connected to the RF Out test port.
CalReady applies to the entire frequency range of the FieldFox. It can be used to
check the integrity of the jumper cable.
Cal ON
is shown on the screen when powered ON or Preset.
When measuring a DUT using a jumper cable or adapter - NOT a direct
connection to the RF Out connector – then a QuickCal or O,S,L Cal is
recommended.
The factory CalReady error terms are updated ONLY when you send your
FieldFox to Agilent for Instrument Calibration. Learn more about Instrument
Calibration on page 205.
QuickCal
QuickCal (option 111) is ideal for making DTF and 1-Port Cable Loss
measurements at the end of a jumper cable or adapter. QuickCal corrects for
phase shift, time delay, and loss, of adapters or jumper cables.
How to perform a QuickCal
Press
Esc
at any time during the QuickCal to stop the calibration.
In CAT, NA, or VVM Mode, press
Cal 5
.
Then (if necessary) press
Cal Type
to see
[QuickCal]
.
Then
Start Cal
.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...