64
N9912A FieldFox User’s Guide
If a jumper cable or adapter is required to connect the DUT to the FieldFox,
then connect
those
components to the FieldFox RF Out connector NOW. The
effects of those components will be measured and removed during the
calibration, and only the effects of the DUT will be displayed in the
measurement results.
These should be high-quality components!
In CAT or NA Mode, press
Cal 5
.
Then (if necessary) press
Cal Type
to choose
O,S,L
.
Then (if necessary) press
Connector Type
.
Then using the
▲|▼
arrows or the rotary knob, select the connector type and
gender of your DUT port to be measured and press
Enter
. This will determine
the type and gender of cal kit to use. For example, if a Type-N Male connector
is used on the DUT port to be measured, then the Type-N Male standards
(OPEN, SHORT, AND LOAD) will ALL be connected in place of the DUT during
the calibration.
If there is more than one Cal Kit for the selected connector type, press
Cal Kit
and select the correct Cal Kit to use.
NOTE
Calibration kit definitions are unique to each connector type and connector sex.
They must be correctly applied to obtain the most accurate calibration.
Then press
Start Cal
.
Press
Open
,
Short
, and
Load
in any order.
Connect the corresponding standard at the point where the DUT will be
connected, then press
Measure
.
Measure each of the three standards. Standards can be measured again at any
time if you feel a standard was not connected securely.
Press
Finish
to complete the O,S,L calibration.
Normalize
Normalize is used to calibrate a
2-port Insertion Loss
measurement (CAT Mode)
and a
S21 Transmission
measurement (NA Mode). One of these measurements
must be selected before performing a Normalize Cal.
NOTE
A Normalization Cal is REQUIRED to make
accurate
2-port Transmission or
Insertion Loss measurements.
How to perform a Normalization
Press
Cal 5
.
Cal Type [Normalize]
should appear.
Connect a short, high-quality, phase stable cable between the FieldFox
RF OUT
and
RF IN
connectors.
NOTE
All subsequent insertion loss measurements are made relative to the insertion
loss of this cable. For example, if you use a Normalization cable with 1 dB of loss,
then after pressing
Normalize
, the display will show 0 dB of loss with this cable
in place. Therefore, for highest accuracy, when measuring the DUT also attach
the cable that was used in the normalization cal.
Press
Start Cal
then
Thru
then
Measure
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...