Calibration for CAT, NA, and VVM Modes
65
Press
Finish
.
Connect the DUT between the
RF OUT
and
RF IN
connectors.
NOTE
A Normalization cal will NOT be interpolated if the Resolution or Frequency
range changes.
CAL ON U
is shown on the screen when a Normalization Cal is correcting the
measurement.
Interpolation
Highest measurement accuracy is achieved when the measurement type,
frequency range, power level, and resolution settings remain the same during the
measurement as when the FieldFox was calibrated. If these settings change after
a calibration, the FieldFox will interpolate the calibration so that VERY accurate
measurements continue to be made.
Interpolated Calibrations are only
slightly less accurate
than a calibration
performed at the measurement settings. For example, if frequency settings
change after a Preset Calibration is performed, the Preset Calibration is only
slightly less accurate. Learn more about the relative accuracy of FieldFox
calibrations on page 68.
When a calibration is being interpolated, an asterisk is added to the screen
annotation as follows:
Cal ON*
For highest accuracy
: a new cal should be performed when the temperature
changes more than about 10
°
F (5
°
C), or when the connection to the DUT requires
a different jumper cable or adapters.
How to turn Error Correction OFF
“Error Correction” is applying the results of a calibration to a measurement.
Error Correction is automatically turned ON after a calibration. You could turn
Error Correction OFF in order to see the effects of a calibration on the
measurement.
Press
Cal 5
then
Error Correction ON OFF
.
Then
Exit
or
Esc
.
Cal OFF
is shown on the screen when Correction is OFF.
Cal OFF*
is shown on the screen when switching from a 1-port measurement
to a 2-port measurement or vice-versa. Calibration for the current
measurement has NOT been performed.
Cal ON ? – Questionable Accuracy
If a change made by the user to the output power setting, or a change made
automatically by the FieldFox firmware makes accurate calibration interpolation
questionable, a question mark is added to the screen annotation for all
calibration states (Ex:
Cal ON ?
). The resulting measurement accuracy depends
on how much the setting has changed. For optimum accuracy, recalibrate using
the new settings.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...