SA (Spectrum Analyzer) Mode
91
o
Log
Averaging – Best for displaying Trace Averaging.
LgAv
is shown on the
left side of the FieldFox screen when selected.
o
Power
(Linear) Averaging – Best for measuring true power levels. Used in
Detection Average and Noise Marker Average. Mathematically, trace noise is
2.5 dB higher than when using Log Average.
PAvg
is shown on the left side
of the FieldFox screen when selected.
Average Count
The Average Count setting is used mainly with the Average Trace State described
above. In this Trace State, the Average Count setting determines the number of
sweeps to average. The higher the average count, the greater the amount of noise
reduction.
When Trace (display) State is set to
Average, MaxHold,
or
MinHold
, the average
counter is shown in the left edge of the screen below the Average Type.
For all three of these Trace States, when
Sweep 3
Continuous
is set to
OFF
,
press
Restart
to reset the sweep count to 1, perform <n> sweeps, then return to
Hold
.
How to set Average Count
Press
Meas Setup 4
Then
Average Count
Enter a value from 1 to 10,000 using the numeric keypad, the
▲|▼
arrows, or
the rotary knob.
Press
Enter
Detection Method
In SA Mode, the X-axis is comprised of data points, also known as “buckets”. The
number of data points is specified using the Resolution setting. Learn how on
page 89.
Regardless of how many data points are across the X-axis, each data point must
represent what has occurred over some frequency range and time interval.
From the frequency span of the measurement, the span of each data point is
calculated as (frequency span / (data points-1)). The detection method allows
you to choose how the measurements in each bucket are displayed.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...