92
N9912A FieldFox User’s Guide
One bucket showing Positive peak, Sample, and Negative peak Detection methods.
How to set Detection Method
The current Detection method is labeled on the left edge of the screen. When a
method is selected manually, a # precedes the label. For example: #
Nrm
means
that
Normal
was selected from the softkeys.
Press
Trace__6
Then
Detector
Then choose from the following
:
o
Auto
Displays the most appropriate Detection method based on other
settings.
[ # is NOT shown]
o
Normal
[Nrm]
Provides a better visual display of random noise than
Positive peak
and avoids the missed-signal problem of the
Sample
Mode.
Should the signal both rise and fall within the bucket interval, then the
algorithm classifies the signal as noise. An odd-numbered data point
displays the maximum value encountered during its bucket. An even-
numbered data point displays the minimum value encountered during its
bucket. If the signal is NOT classified as noise (does NOT rise and fall) then
Normal is equivalent to Positive Peak.
o
Positive Peak
[Pk]
Displays the maximum value of all the measurements in
each bucket. This setting ensures that no signal is missed. However, it is not
a good representation of the random noise in each bucket.
o
Negative Peak
[NPk]
Displays the minimum value of all the measurements
in each bucket.
o
Sample
[Smp]
Displays the center measurement of all the measurements in
each bucket. This setting gives a good representation of the random noise in
each bucket. However, it does not ensure that all signals are represented.
o
Average
[#RMS]
Displays the Root Mean Squared (RMS) average power of
all the measurements in each bucket. This is the preferred method when
making power measurements. Learn more on page 144.
Display Line
A display line is a simple, horizontal line that can be placed at any amplitude
level on the SA screen. Use a display line as mental guide for visual feedback. A
display line is similar to a Limit Line, except that no PASS/FAIL testing occurs.
A display line is easier to create than a Limit Line. Learn about Limit lines on
page 145.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...