System Settings
199
Use
Hold
to conserve battery power or to allow you to save or analyze the results
of a specific measurement sweep.
The current sweep mode is shown on the screen as:
Continuous
Hold
Hold
-->
is displayed while a single sweep occurs.
How to perform a single sweep while in Hold
Press
Sweep 3
.
Then choose one of the following:
o
Single
Automatically sets Continuous OFF (Hold) and causes FieldFox to
make ONE measurement sweep, then hold for the next Single key press.
o
SA Mode ONLY
– With
Continuous OFF
each successive
Run/Hold
press
also causes a single sweep.
o
Continuous ON OFF
Makes continuous sweeps. This is the typical setting
when battery power is not critical.
Preset
The Preset key restores the FieldFox measurement settings to a known condition.
You can preset the entire instrument, or just the current Mode (CAT, SA, and so
forth).
When Preset, the FieldFox CalReady calibration is recalled for use in CAT and
NA Modes. Learn more on page 65.
In addition to all
Preference
settings, the following settings always survive
Preset and Mode Preset:
External Reference state (page 209)
SA Units (page 86)
SA Correction states (page 89)
VVM Display Resolution (page 168)
Marker Peak Excursion and Threshold (page 182)
How to Preset the FieldFox
Press
Preset
.
Then choose one of the following:
o
Preset
Returns ALL Modes to the factory default settings.
o
Mode Preset
Returns only the current Mode to the factory default settings.
The current settings for all other Modes remain.