10. Connect an unterminated, longer cable to the FieldFox
PORT 1
connector.
11. Observe the displayed phase difference between the reference cable and the
untrimmed cable.
NOTE
The VVM phase reading shows a maximum of ±180°. Therefore, the electrical
length of the untrimmed cable MUST be within 180° of the reference cable.
12. Carefully trim the cable until the phase shift reads zero. The attached cable’s
electrical length is now matched to the reference cable.
13. Repeat steps 10 through 12 for the remaining cables to be trimmed.
2-Port Transmission Measurements
A 2-Port Transmission measurement is used for measuring electrical length,
insertion loss, gain, or isolation of a DUT at a single CW frequency. The FieldFox
signal source is transmitted out the port 1 connector, through the DUT, and into
the port 2 connector. In the following image, the gain of an amplifier is being
measured.
2-Port Transmission measurement of an amplifier
How to make a 2-Port Transmission Measurement
1. Press
Mode
then
VVM
then
2-Port Transmission
.
2. Press
Freq/Dist
and enter the measurement frequency.
3. Press
Meas Setup 4
then
More
then
Output Power
. Select
High
for
passive devices. Select
Low
for amplifiers.
4. Calibrate the FieldFox using one of the methods described on page 170.
Before performing a QuickCal or Mechanical Cal, press
Meas Setup 4
then
Zero OFF
.
5. Connect the DUT.
Notes:
The magnitude value is the gain (positive value) or the loss (negative value) of
the DUT.
The phase value is the difference in phase (in degrees) between the DUT input
and output.
To measure isolation of the amplifier, reverse the connection to the amplifier
(
PORT 1
to the amplifier output).