System Settings
211
2.
Voltage
setting from 1.00 V to 32.00 V with .01 V resolution.
3.
Current:
The amount of current (in Amperes) being drawn from the source.
4.
Max Current:
A conservative estimate of maximum current possible given
the present voltage setting. The actual ‘trip’ level may be slightly different.
How to make Voltage Source settings
Press
System 7
Then
System Configuration
.
Then
Voltage Source
.
Then
Source ON OFF
to enable and disable the voltage source output.
Then
Voltage x.xx V
Using
▲|▼
arrows, the rotary knob, or numeric keypad,
enter a value between .001 and 32 volts.
Clear Overload
If the voltage source becomes ‘TRIPPED’ press to re-enable
the voltage source output. Then reset
Source ON.
Status Line ON OFF
Turn ON and OFF the voltage source status line that
appears near the top of the screen. The status line is always ON while making
Voltage Source settings.
Security Level
For security reasons, you can prevent frequency information from appearing on
the FieldFox while in SA, NA, or CAT modes.
How to set Security Level
Press
System 7
Then
System Configuration
Then
Security Level
Then choose from the following:
o
None
All frequency settings are visible.
o
High
Frequency information is blanked from the following:
o
Display annotation
o
Softkeys
o
Marker display and marker table
o
Calibration properties
o
All settings tables
o
Limit line tables
o
All saved .png files
o
Any of the following will re-display frequency information:
o
Set to
None
, Preset, Mode Preset, or FieldFox restart.