124
4
Hold the paper clip in place, press the gripper
release button and rotate the gripper arm to
the right.
5
The gripper arm will come off.
6
Replace the gripper arm by holding the paper
clip in place, pushing the gripper arm into the
holder and rotating the gripper arm to the left.
On completion of this procedure:
❏
Install the front cover.
❏
Turn the power to the autosampler on.