98
Troubleshooting and Test Functions
Jittery (shaky) movement in X and or theta axes and/or when the needle goes
through the gripper arm into the via
Step 4
Check the tension of the belts.
For this use the
Macro2.mac
and measure the torque for theta and X-axis.
•
If the theta or X torque is not in the range, proceed with STEP 5 (if you
think you can adjust the torque). Otherwise proceed with STEP 7.
Step 5
Adjust the belts tension.
•
If the measured torque value is too low, the belt needs to be tightened.
•
If the measured torque value is too high, the belt needs to be loosened.
For this, slide the motor (X or theta) on the holder bracket in the appropriate
direction and test the tension with the
Torque2.mac
macro. Repeat this
steps until the values are in the appropriate torque range.
Step 6
Reset the sampler alignments to default value.
Reset tray alignment, and transport alignment is possible with the Control
Module and the Chemstation. To reset the transport alignment with the
Chemstation enter following command in the command line.
Print sendmodule$(lals, “tray:alig 0.00,0.00”)
Step7
Exchange the sample transport assembly (part number G1313-60009
or G1329-60009).
Step 8
Exchange the main board (part number G1313-69520 or
G1329-69520).
Typical ranges
Theta
(both)
30-50
X-axis (both) 50-90
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......