104
1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Needle centering over the vial or the well
Action to take if the needle does not hit in the safe area
✔
Check if the right vials or plates are used (see) or (see
✔
Make sure the needle is correctly installed. It should be pushed into the
needle carrier as far forward as possible and centered in the vial pusher.
✔
Upgrade the firmware revision to A.04.14 or higher and the ChemStation
revision to A.08.04 or higher
✔
Perform an auto-alignment (without any plates in place)
✔
Change the tray G1367-60001 (see service note G1367-007)
N O T E
The positioning of the needle is very precise. You have to take no action if the needle hits in
the safe area.
Figure 16
Vial cap
metal border
Septum
Safe area to hit
N O T E
If the diameter for the safe area is approximately
1mm smaller
than the diameter of the
septum no corrective action is necessary.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......