124
1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Removing the Loop Capillary
1
In the user interface start the maintenance mode and select the
“ChangeLoop Capillary” function.
2
Open the front door and remove the side door.
3
Remove the plate tray from the tray base.
4
Push the silicon safety tube over the needle
5
Remove the corrugated loop capillary waste tube and introduce the small
tubing from the leak kit into the loop capillary protection tube.
6
Draw up the liquid with the syringe.
7
Unlock the needle tighter lock system.
8
Draw up the rest of the solvent from the loop capillary protection tube.
9
Loosen the loop capillary fitting on the analytical head side.
When required
Capillary blocked
Capillary broken
Tools required
Two 1/4 inch-5/16 inch wrenches 8710-0510 (supplied in accessory kit)
Parts required
G1367-87300 Loop capillary (injection volume up to 100 µl)
for the G1367A/68A
G1375-87315 Loop capillary (injection volume up to 8 µl)
for the G1377A/78A
G1377-87300 Loop capillary
(injection volume up to 40 µl)
for the G1377A/78A
WA R N I N G
When opening capillary or tube fittings, solvents may leak out. Please observe
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
N O T E
If the loop capillary is not broken and no solvent has leaked into the loop capillary tube, the
solvent draw up steps using the syringe can be skipped (steps 5, 6, 8).
C A U T I O N
An uncovered needle is a risk of harm to the operator.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......